The LSProbe 1.2 Field Probe is a high speed, high accuracy and high dynamic range electric-field probe. Its standard frequency range of 9 kHz – 8.2 GHz.<br>Best-in-class compensation of linearity, frequency and temperature guarantees accurate measurements from less than 0.1 to at least 1,000 V/m. A dynamic range of 100 dB is achieved for many frequencies, enabling field measurements at more than 10,000 V/m
A single axis, continuously sampling mode, operating at 2 MS/s, can be used for Equivalent Isotropically Radiated Power (EIRP) measurements of IoT products without antenna connectors, in accordance with EN 300 328 and EN 301 893.
You can also read out the measurement data of all three axes simultaneously with a snap of your fingers.
Laser-powered operation eliminates battery recharging and replacement. Extensive factory calibration data are provided with each field probe and is handled automatically by the LUMILOOP Software (TCP Server and GUI).
LUMILOOP is an DAkkS certified Accredited Calibration Lab for E-Field Strengths. We deliver accredited calibrations up to 26.5 GHz and up to 600 V/m, ensuring reliable performance for demanding EMC, RF, and industrial applications.
Test Automation Integration
- Rohde & Schwarz – ELEKTRA
- Rohde & Schwarz – EMC32 – Version 10.0
- Nexio – BAT-EMC
- Ametek – Compliance 5
- Ametek – Compliance and Immunity 6 (CIS 6)
- ETS-Lindgren – TILE!
- Raditeq – RadiMation
- Teseq – WIN 6000
- Techno Science Japan – TEPTO
- AR – Emcware
- Frankonia – ProveEMC
- TDK RF Solutions – Radiated Immunity Lab
- Toyo – VI5RS – Radiated Immunity for Automotive testing
- Toyo – IM5RS – Radiated Immunity for IEC 61000-4-3 testing
Variants of LSProbe 1.2
| Variant | Frequency | Field Strength | Application | Computer Interface Type | Part No. |
|---|---|---|---|---|---|
| E | 9 kHz to 8.2 GHz | <1 V/m … >1.000 V/m | – Universal radiated immunity – Automotive & military radar pulses – Reverb chambers | CI-250 System CI-250+ System | 1111 1115 |
| F | 10 Hz to 8.2 GHz | <3 V/m to >1.000 V/m | – High voltage applications – Line frequency (e.g. 16⅔, 50/60 Hz) | CI-250 System CI-250+ System | 1112 1116 |
| G | 9 kHz to 8.2 GHz | <15 V/m to >15.000 V/m | – High field strength signals & pulses | CI-250 System CI-250+ System | 1113 1117 |
LSProbe 1.2 CI-250 System / LSProbe 1.2 CI-250+ System
Scope of Delivery – LSProbe 1.2 CI-250+ System
The LUMILOOP LSProbe 1.2 Field Probe System consists of an LSProbe 1.2 E-Field Probe (Variant E, F, G by choice) with a fixed length of fiber optic cable, a Computer Interface CI-250+ / CI-250. Optional accessories—such as probe stands, a fiber-connector cleaning kit to enable repeatable and efficient measurement setups, or outdoor fiber-optic cables—may complement and complete your laboratory equipment.
The scope of delivery from LSProbe 1.2 CI-250+ System and LSProbe 1.2 CI-250 System may vary.
| LSProbe 1.2 CI-250+ System E-Variant, Part No. 1115 (exemplary) |
|---|
| 1x 1011, LSProbe 1.2 Field Probe Variant E – permanently attached 5 m optical fiber cable – frequency range & factory calibration (non-accredited): 9 kHz to 8.2 GHz – field-strength range 9 kHz to 30 MHz: <1 V/m to >10 kV/m – field-strength range 30 MHz to 4 GHz: <0.1 V/m to >1.5 kV/m – field-strength range 4 GHz to 6 GHz: <0.5 V/m to >1 kV/m – field-strength range 6 GHz to 8.2 GHz: <3 V/m to >1 kV/m – sampling rate of 500 kS/s in continuous mode, 2 MS/s in burst mode 1x 1304, Optical Fiber Extension Cable 15m 3x 1314, Sacrificial Cable Kit with E2000 couplers 1x 1333, 1/4-20 UNC 16 mm plastic screw 1x 1006, Computer Interface CI-250+ – Computer Interface with 4.3″ Touchscreen and Ethernet interface – 1,000 mW max. laser output power, laser class 1M 1x IEC 60320 C13 cable 1x 1328, LSProbe CI-250+ System Carrying Case 1x 1004, LUMILOOP Software Package on USB flash drive – incl. factory calibration data and accredited calibration factors if applicable |
Options:



We Take Care of Your Calibration!
Our all-round carefree service for LSProbe E‑Field Probes and LSPM RF Power Meters includes checking your measuring equipment and performing the ISO 17025 accredited calibration.
Click here to learn more about our Electric-Field Calibration and Measurement Capabilities (CMC) and the LUMILOOP Calibration Procedure.
LUMILOOP provides Rack Mount Kits that enable seamless laboratory integration and offer an efficient solution for maintaining a clean and organized measurement environment. For further dertails, please refer the LSFrame / Rack Mount Kits page.
| 19″ Rack Mount Kit for 1x CI-250+ | Part. No. 3026 |
| 19″ Front Panel, 2x Computer Interface CI-250 | Part. No. 3022 |
LSProbe Accessories
Please check the LUMILOOP accessories page for further information on LSProbe stands, cables and fiber cleaning devices.
Stands for E-Field Probes
The table top probe stand base allows for mounting the pole for positioning the field probe.
Part No. 1307
The table top probe stands are available for positioning the LSProbe center above the table in the following heights: 100 mm, 125 mm, 150 mm, 200 mm, 300 mm.
The Mounting poles have a relative permittivity better than 2.7 at 1 kHz.
| Height above table | Part No. |
|---|---|
| 100 mm | 1308 |
| 125 mm | 1309 |
| 150 mm | 1310 |
| 200 mm | 1311 |
| 300 mm | 1312 |
LSProbe Table Probe Stand, 77 mm, 90° angle
Base Plate (Part No. 1307) + Mounting Pole, 77 mm, 90° (Part No. 1326)
Ideal for 154 mm TEM Cell measurements, mounting pole for positioning the Field Probe’s center 77mm above surface, in 90° angle. It allows for quick mount and release as well as easy alignment of LSProbe Field Probe and has a relative permittivity better than 2.7 @ 1 kHz
The flexible tripod probe-stand for versatile positioning of LSProbes allows for quick mount and release for vertical positioning approximately 150 mm to 220 mm above surface.
A strong magnetic feet with rubber coating holds the LSProbe in position.
Part No. 1306
The lightweight floor probe stand is metal-free, has very low absorption, and a height of 2.45 m above ground.
The probe can be adjusted variable at any angle in the height needed from 0.2 m – 2.53 m.
A ruler shows the exact LSProbe height above ground. The floor probe stand has lockable swivel castors and base features slots for repeated alignment.
The Floor Probe stand is equipped with a single probe holder. Additional Probe Mounting Bracket are available on demand.
Additionally, the floor probe stand is optimized for IEC 61000-4-3 16 point field homogeneity validation.
Part No. 1316
Probe mount for floor probe stand for 1 additional probe.
Part No. 1317
LSMove Field Probe Positioning System
The LSMove Positioning System is a companion product to the LSProbe Laser-Powered E-Field Probe, enabling automated measurements of field distributions in 3D testing volumes. LSMove’s unique design minimizes the field distortion introduced by the positioning system, enabling high-accuracy measurements.
LSMove accelerates:
LSMove is designed for high positioning accuracy and repeatability, requiring no zeroing upon start-up. Its non-conductive probe mount, probe suspension and side stands integrate seamlessly with LUMILOOP’s LSProbe E-Field Probes. Fiber-optic cables for the Field Probe are embedded in the straps for hassle-free positioning.
LSMove is connected via USB and includes software for manual positioning and automated operation via TCP/IP using standard SCPI commands.
LSMove speeds up:
- IEC 61000-4-3 anechoic chamber calibration
- Antenna pattern characterization
- IEC 61000-4-20 TEM-waveguide calibration, and
- IEC 61000-4-21 reverberation chamber calibration
Additional features include:
- Long-time operation
- Quick swap, fast-charging battery pack
- Fiber-optical control ports
- Wireless remote control
- Quick and simple position calibration
- Built-in hazard zone avoidance
Fiber Optic Cables
All LSProbe / Laser-Powered RF Power Meter System cases are equipped with the necessary products to plug-and-measure right when receiving the delivery.
LUMILOOP offers additional accessories.

A set of two 0.5m duplex fiber optic cables with FC/ST and E2000 connectors. The automatic closures of the E2000 connectors prevent contamination of the connection surfaces.
| Article | Part No. |
|---|---|
| Sacrifical Cable Kit | 1314 |
| E2000 Coupler for Sacrificial Cable Kit | 1320 |
We offer a sacrificial cable kit with customized length. This is a sacrificial cable with 6m length, a sacrificial cable with 0.5m length, duplex E2000 coupler and one pair of FC/ST couplers.
Part No. 1351
A separate fiber optic extension can be supplied on request in the following lengths: 5m, 10m, 15m, 20m, 30m, 50m. The cable is supplied with FC/ST connectors and a separate pair of FC/ST couplers . The LSProbe 1.2 allows for use with 1.000m optical fiber extension. The LSProbe 2.0 allows for use with 500m optical fiber extension.
We also offer outdoor fiber optic extensions in 30m and 50m lengths. On request we can also supply dual fiber optic extensions.
| Cable length | Part No. |
|---|---|
| Optical Fiber Extension 5m | 1302 |
| Optical Fiber Extension 10m | 1303 |
| Optical Fiber Extension 15m | 1304 |
| Optical Fiber Extension 20m | 1305 |
| Optical Fiber Extension 30m | 1321 |
| Optical Fiber Extension 50m | 1322 |
| Optical Fiber Extension 100m | 1323 |
An outdoor optical fiber extension is available in the following lengths: 20m, 25m, 30m, 50m, 100m, 200m.
The duplex optical-fiber extension cable is supplied with FC/ST connectors and a separate pair of FC/ST couplers.
On request we also supply dual fiber optic extensions.
| Cable length | Part No. |
|---|---|
| Outdoor Optical Fiber Extension 20m | 1334 |
| Outdoor Optical Fiber Extension 25m | 1339 |
| Outdoor Optical Fiber Extension 30m | 1335 |
| Outdoor Optical Fiber Extension 50m | 1336 |
| Outdoor Optical Fiber Extension 100m | 1337 |
| Outdoor Optical Fiber Extension 200m | 1338 |
LUMILOOP offers robust cable drums for duplex fiber optic cable. The drum is supplied with a sacrificial cable with an E2000 connector, which is connected to the extension cable, the E2000 coupler is integrated into the cable drum. The cable reel is suitable for indoor and outdoor fiber optic cables (these are not included in the scope of delivery). The small cable drum is suitable for fiber optics up to 100m. For fiber optic lengths over 100m, we offer large cable drums.
On request we also offer dual cable drums.
| Article | Part No. |
|---|---|
| Cable Drum Small <100m | 1331 |
| Cable Drum Large >100m | 1332 |
Fiber Connector Cleaning

The fiber connector cleaning kit allows for cleansing the fiber optic connector to ensure reliable operation of the LSProbes.
The set consists of 400x optical fiber microscope, two packs of lint-free cassette cleaning-wipes, fiber optic cleaning-solution pen, an unfilled isopropyl alcohol (IPA) pipette/bottle and spare FC/ST dust caps and two E2000 locking caps.
For further information please read the application note 4 – AN#4 Reliable Operation of LSProbes Laser-Powered Electric-Field Probes.
Further adapters for FSMA connectors for the fiber microscope are available on request.
PartNo. 1319
LSProbe 1.2 Specifications
For detailed technical specifications, please take a look at the LSProbe 1.2 datasheet.
| Frequency Range | 9 kHz – 8.2 GHz (Please check differences for variants E, F, G) |
|---|
| Damage Level | >25 kV/m |
|---|
| Resolution | <0.01 dB |
|---|
| Amplitude Accuracy | Accredited Cal. at LUMILOOP |
|---|---|
| 9 kHz … 1 GHz | ±1.2 dB |
| 1 GHz … 8.2 GHz | ±1.5 dB |
| Typical Worst-Case Isotropy Error | |
|---|---|
| @ 1 GHz | ±1.0 dB |
| @ 3 GHz | ±1.7 dB |
| @ 6 GHz | Please see Application Note 8. AN#8 – Measuring Isotropy from 1 to 6 GHz |
| Linearity Error | ±0.2 dB relative to 10 V/m |
|---|
| Temperature Stability | ±0.1 dB |
|---|
| Ambient Temperature | -5 °C … 40 °C |
|---|