Lumiloop support you in E-Field uniformity and homogeneity testing acc. IEC 61000-4-3 E-Field.
Please read the following measurement set-up suggestions.

1x LSProbe + LSPM

The combination of one LSProbe laser-powered E-field probe and one LSPM RF Power Meter is the easiest way to get IEC 61000-4-3 field uniformity measured fast.

You can expect the following measurement times: 10-20 seconds per point

System Configuration Options for 6 GHz / 18 GHz:

1. Basic System, USB connection

2xLSProbe CI-250 with LSPM 1.0 in 19″ Rack

2. CI-250+ System

CI-250+ with LSPM 1.0 in 19″ Rack

Recommended accessories:

  • LSProbe Floor Probe Stand (part no. 1316)
  • Fiber Connector Cleaning Kit (part no. 1319)
  • Fiber Cable Extensions as needed

4x LSProbe + LSPM

You have four LSProbes anyway, e.g. for ISO11451-2?
Combine a 4x Multi-Probe System (LINK) with a LSPM RF Power Meter to measure one column, 4 of the 16 positions, at once!
Our Floor Probe Stand makes positioning the probes easy.

System Configuration Options for 6 GHz / 18 GHz:

1. Basic System, USB connection

4xLSProbe CI-250 with LSPM 1.0 in 19″ Rack

2. CI-250+ System

1x CI-250+, 3xLSProbe CI-250 with LSPM 1.0 in 19″ Rack

Recommended accessories:

  • LSProbe Floor Probe Stand (part no. 1316) with three additional probe mounts (part no. 1317)
  • Fiber Connector Cleaning Kit (part no. 1319)
  • Fiber Cable Extensions (…)

LSMove + 1x LSProbe + LSPM

LSMove, our automated probe positioning system, elevates field uniformity measurement to the next level.
LSMove eliminates manual probe positioning completely, and creates another reduction in measurement time.
Once the system is set up, 16 points can be measured in no time!

However, there is a stronger motivation for automated positioning.
Have you ever wondered how the standard 50 cm grid defined in IEC 61000-4-3 comes together with a wavelength of 5 cm at 6 GHz?
Would you like to SEE where the best uniform area is located, and where the field uniformity is not ideal?
With the combination of LSMove, LSProbe and LSPM, you can easily measure hundreds of positions in one day!

E-Field homogeneity distribution in chamber